Emerson Process Management ROSEMOUNT 5300 Manuale Utente Pagina 158

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 324
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 157
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
7-28
REMOVING THE
TRANSMITTER HEAD
1. Loosen the nut that connects the
transmitter housing to the Process Seal.
2. Carefully lift the transmitter head.
3. Make sure the upper surface of the
Process Seal is clean and the
spring-loaded pin at the center of the
Process Seal is properly inserted (the pin
should move back when pushed into the
hole).
4. Attach the protection plug to the Process
Seal.
Nut
Process Seal
Put the protection
plug here!
FLANGE VERSION
NOTE!
Do not remove the Process Seal from the
adapter!
THREADED VERSION
Put the protection
plug here!
Nut
Process Seal
Adapter
Vedere la pagina 157
1 2 ... 153 154 155 156 157 158 159 160 161 162 163 ... 323 324

Commenti su questo manuale

Nessun commento